Ultra-Low Temperature Freezers

Series Overview Control System: Digital temperature control. Refrigeration System – CFC Free: Free: Cascade Type Two hermetic type compressors Large air cooled Condenser – Cooling coils Copper tubing Download Specifications

SKU# UL

Share this Product

About

Freezer has an automatic timer system that will restart the unit in case of power failure.

Frost and ice build-up along the gasket is minimized by a heater harness located beneath the channel cap away from the inner chamber.

Digital temperature control with dual display and adjustable temperature range. Scan button for setting temperature and tamper proof lockout.

Quick pull-down to operating temperature with efficient and reliable cascade refrigeration. Two large hermetic compressors ensure rapid pull-down of product load. Because of the unique design of the freezers, the pressure in the refrigeration system is held very low, eliminating excessive heat build up. This low pressure allows the freezer to operate efficiently, extending compressor life.

Compressor life is extended by an energy-saving low stage which operates only on demand.

Compressors are protected with a constant flow of refrigerant.

Heat is efficiently and effectively dissipated by an air-cooled condenser with two heavy-duty fans. Condenser requires no liquid coolant.

Additional information

Control System

Digital temperature control

Refrigeration System - CFC Free

Cascade Type –
Two hermetic type compressors – Large air cooled Condenser – Cooling coils – Copper tubing

Insulation - CFC Free

Minimum of five inches of polyurethane – Sublids – Double sealed gaskets

Construction: Inner Chamber

– Heavy gauge corrosion resistant zinc coated galvanized steel.
Outer shell – heavy gauge, cold-rolled steel

Electrical Requirements:

115 volts, 60 hertz, 1 phase
208 volts, 60 hertz, 1 phase
230 volts, 60 hertz, 1 phase
220 volts, 50 hertz, 1 phase
440 volts, 60 hertz, 1 phase

Digital Alarm System:

Battery operated alarm system with digital display. Audio and visual display will activate with temperature warm up or loss of power. System includes an audio silence switch and alarm test switch.

Accessories

Chart recorder 7 day 6 inch
Inventory control systems
Porthole access
CO2 and LN2 back-up system
Stainless steel chambers
Racks – Baskets
Cold safety gloves
Three phase electrical wiring
Telephone alarm system
Voltage safety system
RS232 or RS485 Data Ports