Stability Chamber – PH055

The PH series chambers are designed for performance specifications exceeding FDA and ICH requirements. For GMP studies, ultrasonic humidification provides excellent humidity control and avoids hot spots during mapping. Thermoelectric cooling provides more stable temperature control than equivalent refrigeration systems. All PH series chambers incorporate inherent cooling redundancy, quiet operation, simple serviceability and excellent reliability. The PH055 is a 2-door stability chamber Popular options include chart recorder and TraceLock electronic/audit capable locks. Download Specification

SKU#

Share this Product

About

The PH series was introduced in 2003 as the first commercially manufactured large capacity Peltier-cooled pharmaceutical stability testing chamber. Over a thousand PH series stability chambers are presently in use by pharmaceutical, cosmetic, nutraceutical, medical device, and research companies around the world. The advantages of these chambers for stability testing include:

Cooling Redundancy

Each PH series chamber includes multiple, independent thermoelectric cooling units. The 7 year warranty on cooling components and the ability of our chambers to maintain temperature even if the chamber suffers multiple failures is unmatched by any other refrigeration-based chamber.

Simple Serviceability

Due to the omission of all refrigeration equipment on these models, refrigeration technicians are not required during chamber servicing. Refrigerated stability chambers are not only costly to service, but can require days to service properly. Nearly every component of thermoelectric-based chambers can be serviced within minutes with basic tools. For instance, the ultrasonic humidifier is capable of being removed and reinstalled in less than a minute and operates on non-hazardous 24 volts.

Superior Control

The PH series chambers control well within the specifications required by the FDA and ICH for GMP studies. The ultrasonic humidification system provides excellent humidity control and avoids hot spots seen during chamber mapping of steam boiler equipped chambers. Thermoelectric cooling and automatic switching system from cooling to heating control provide consistent results throughout the available temperature range. Standard control at the sensor in these chambers is ±0.2°C and ±0.3%RH.

Reduced Energy Loads

The PH Series chambers utilize less energy than comparable stability chambers and introduce less impact upon building cooling systems. This efficiency “dual benefit” allows our chambers to be located in spaces unsuitable for other chambers. Electrical energy savings by the PH series vs. steam boiler equipped chambers can easily amount to hundreds or even thousands of dollars saved per year.

Proven Controllers

Standard controllers for the PH series are manufactured by Fuji Electric and are ideal for stability testing chambers. Unlike many proprietary controllers, this controller is commercially available and proven in tens of thousands of installations. Standard functions include: autotuning, fuzzy logic, PID control, programmable alarms, calibration capability, offset capability, etc. A touchscreen control interface is optional. Other controller manufacturers are also supported (Watlow, Allen Bradley etc.)

Service and Warranties

We offer 24 hour technical support throughout the lifetime of your chamber. In addition to our standard warranties, we also offer extended warranties for cooling components, parts and labor. We also provide qualifications, validations and preventive maintenance services at an additional cost. Please ask for a quote.

Options Available

Chart Recorder, Data Loggers, Dryers

Quiet Operation

The PH chambers utilize quiet and highly efficient DC fans. Locating these chambers in laboratory workspaces is a viable option.

Additional information

Cabinet Construction

Stainless model has stainless steel front door, and sides on exterior. Interior is stainless steel side walls, back, floor, ceiling, and door liner. This unit includes door switch controlled LED interior lighting.

Standard model features stainless steel door, stainless steel front, and brushed aluminum sides. Interior has anodized aluminum side walls, door liner, and back with stainless steel floor and ceiling.

2" O.D access port with access port cover and insulation plug located on the side of the chamber.

Doors

Positive seal, self-closing doors (gravity cam-lift) with 120 degree stay open feature. Door gasket removal and installation is tool-less. Glass door models feature double pane,”Low-E” glass.

Insulation

Entire cabinet structure and solid doors are foamed-in-place using a high density, polyurethane insulation that has zero ozone depletion potential (ODP) and zero global warming potential (GWP).

Shelving

Wire shelves (3 per door) available in stainless steel or PVC coated. Shelves capable of supporting up to 150 lbs of evenly distributed weight. Gray PVC coated shelves are standard.

Full height shelf support pilasters, with double oblong holes on 1/2″ centers. Pilasters made of the same material as cabinet interior. Stainless steel shelf clips to be included for each shelf.

Cooling System

Top mounted, thermoelectric cooling system. Peltier based technology is solid state, highly efficient, reliable, and provides minimal vibration.

Semi-Redundant cooling capability.

Cooling surfaces act as dehumidifying modules at 25°C/60%RH and 30°C/65%RH

Humidity Control

NIST traceable 0.8% Rotronic humidity sensor w/cert. Solid state controlled, ultrasonic humidifier with ceramic/titanium disk for long life.

Electrical Characteristics

Unit completely pre-wired at factory, and ready for final connection to a 120V/60Hz/1/15a phase – 15 amp dedicated circuit.

Schema