Airflow Pattern
Cabinet Construction
- Standard model features powder-coated door, front, and sides. Interior has 304 stainless steel on door liner, top, bottom, back and sides.
- Two, 2" I.D access ports (one on each side) with access port cover and insulation plug located on side wall of the chamber.
Doors
- Positive magnetic seal. Reversible.
Insulation
- Entire cabinet structure and doors insulated using 2" urethane foam.
Downloads
Download Specification DrawingShelving
- 3 stainless steel wire shelves are standard. Shelves capable of supporting up to 150 lbs of evenly distributed weight.
- Full height shelf support pilasters, with double oblong holes on 1/2" centers. Pilasters made of the same material as cabinet interior. Stainless steel shelf clips to be included for each shelf.
Emergency Shutdown
- Unit is supplied with shut down capability upon attaining set temperature
- Prevents runaway temperature from unintended eradication. Can be bypassed for intended decontamination.
Cooling System
- Top mounted, thermoelectric cooling system. Peltier based technology is solid state, highly efficient, reliable, and provides minimal vibration.
- Semi-Redundant cooling capability.
- Cooling surfaces can be made to act as dehumidifying modules at 25°C/60%RH and 30°C/65%RH.
Popular options include humidity control, lighting, timed lighting, chart recorder and TraceLock electronic/audit capable locks.