CABINET CONSTRUCTION:
• Standard model (pictured at right) features powder-coated door, front, and sides. Interior has 304 stainless steel on door liner, top, bottom, back and sides.
RACKING:
•Racking (as shown in pictures on right) available to to stack units.
DOORS:
• Positive magnetic seal. Reversible.
INSULATION:
• Entire cabinet structure and doors insulated using 2" urethane foam.
SHELVING:
• 2 stainless steel wire shelves are standard. Shelves capable of supporting up to 100 lbs.
• Full height shelf support pilasters, with double oblong holes on 1" centers. Pilasters made of stainless steel. Four (4) chrome plated, stainless steel shelf clips to be included for each shelf.
| Standard Specs | |
| Temperature Range | 10°C to 60°C |
| Temperature Control | +0.2°C |
| Temperature Uniformity | +0.7°C |
| Ambient Temp. Recommendation | 21°+3°C |
| Control Sensitivity | 0.1°C |
| Control Readout | Actual and Set-Point Values |
| Sensor | NIST-Traceable Platinum RTD |
| Controller | Microprocessor based PID |
| Calibration Capability | Yes, actual and set values |
| Alarm | Audible and visible |
| Alarm Type | Deviation in 0.1°C or 0.1%RH |
| Remote Monitoring | Available |
| Electrical | 115v, 60 Hz, 3.9 RLA |
| Interior Cubic Feet | 9 net |
| Required Clearance | 1" all sides |
| Number of Shelves | 2 |
| Cord Length | 7' |
| Crated Weight | 340 lb |
COOLING SYSTEM:
• Top mounted, thermoelectric cooling system. Peltier based technology is solid state, highly efficient, reliable, and provides minimal vibration.
• Semi-Redundant cooling capability.
• Cooling surfaces act as dehumidifying modules at 25°C/60%RH and 30°C/65%RH
ELECTRICAL CHARACTERISTICS:
• Unit completely pre-wired at factory, and ready for final connection to a 115/60/1 phase - 15 amp dedicated (recommended) outlet. • A cord and plug set is included.
• A sole use circuit is recommended.
Popular options include humidity control, lighting, timed lighting, chart recorder, and TraceLock electronic/audit capable locks.