DB052 Laboratory Incubator


DB053 IncubatorCall for pricing 314-534-3111
To print, click the DB052 Incubator spec sheet (pdf).

CABINET CONSTRUCTION:
• 300 series stainless steel front grill, shroud and doors. Anodized quality aluminum exterior ends and back.
• Interior – INSF approved, white vinyl coated aluminum and 300 series stainless floor with coved corners.
• Barrel lock standard for each door. Locks keyed alike.
• Cabinet mounted on a welded, galvanized frame rail, painted for additional corrosion protection.
• Frame rail fitted with 4" diameter casters, (front casters include brakes).

DOORS:
• 300 series stainless steel exterior with white aluminum interior to match cabinet liner. Doors extend full width of cabinet shell. Door gaskets of one piece construction.
• Lifetime guaranteed adjustable torsion type closure system with aluminum block supports. The face of each door fitted with a 12" long recessed handle with sheet metal interlock. Handles are foamed-in-place to insure permanent attachment.

INSULATION:
• Entire cabinet structure and doors insulated using foamed-in-place CFC free polyurethane foam.

SHELVING:
• Adjustable vinyl coated wire shelves (3 per door). Shelves capable of supporting up to 250 lbs.
• Full height shelf support pilasters, with double oblong holes on 1/2" centers. Pilasters made of the same material as cabinet interior. Four (4) chrome plated, stainless steel shelf clips to be included for each shelf.

LIGHTING:
• Incandescent interior lighting. Light(s) activated by rocker switch mounted above door(s).

REFRIGERATION SYSTEM:
• Unsurpassed temperature control through continuous refrigeration bypass system.
• Refrigeration system factory balanced using CFC free 134A refrigerant.
• Top mounted condensing unit.
• Factory pre-engineered capillary tube system with oversized condenser and evaporator to reduce product pull down time and to consistently maintain interior cabinet temperature (non-adjustable capillary tube system superior to problematic expansion valve).
• Evaporator - epoxy coated at the time of manufacture to eliminate the potential of corrosion.

ELECTRICAL CHARACTERISTICS:
• Unit completely pre-wired at factory, and ready for final connection to a 115/60/1 phase - 15 amp dedicated outlet. A cord and plug set to be included.
• A sole use circuit is required.

Specs  
Temperature Range 4°C-60°C
Temperature Control +0.2°C
Temperature Uniformity +1.0°C
Ambient Temp. Tolerance 17°C – 30°C
Control Sensitivity 0.1°C
Control Readout Actual and Set-Point Values
Sensor NIST-Traceable Platinum RTD
Controller Microprocessor based PID
Calibration Capability Yes, actual and set values
Alarm Audible and visible
Alarm Type Absolute Hi/Lo in 0.1°C increments
Remote Monitoring Available
Electrical Requirements 115v, 60 Hz, 12 amps
Interior Cubic Feet 52
Required Clearance 1" all sides
Number of Shelves 6
Cord Length 9'
Crated Weight 420 lbs

DB052 Incubator Schematic